Product Features

:: Multi-Point Push-button Calibration (up to 5 points) with ±1% full scale accuracy
   
:: Automatic or Manual Calibration - for quick, easy calibration; use automatic mode with preset calibration points at most popular values or manually set values in the manual mode for more flexibility
   
:: Auto-Ranging for Conductivity/TDS measurements for fast response and best resolution over wide measurement range
   
:: Adjustable TDS Factor for more accurate measurement (for TDS 6 model only)
   
:: Automatic Temperature Compensation (ATC) for the optimum accuracy under fluctuating temperatures
   
:: Hold Function momentarily freezes reading for easy viewing
   
:: Auto-Power Off saves battery power after non-use
   
:: Self-Diagnostic with message codes for easy troubleshooting
   
:: Large Custom LCD provides optimum viewing even at a distance
   
:: Electrode with Built-in ATC designed for minimal air bubble entrapment during measurement
   
:: Protective Rubber Boot shields meter from drops and features a sturdy built-in stand for easy bench-top operation
   
:: Meter Kit (Optional) - contains everything you need for calibration and measurement in a rugged carrying case


EcoScan
Carrying Kit

Applications

Routine Testing: For quick, accurate conductivity or TDS checks in laboratories, field and schools.

Environmental/Agricultural: Useful in nutrient and fertilizer checks in hydroponics and agricultural industries.

Water Quality Testing: For analyzing water hard water, untreated water, industrial and rinse water, drinking water, effluent water, pool water and incoming process water. Ideal for all types of quality assurance, printing industries and water quality testing.

Specifications

Product Specification

Description
Conductivity Range Not Available
Resolution & Accuracy -
TDS Range 0 to 9.99, 99.9, 999 ppm; 9.99, 99.9 ppt,
|Max. 199.9 ppt (depending on factor setting)
Resolution & Accuracy 0.01, 0.1, 1 ppm & 0.01, 0.1, 1 ppt;
±1% Full Scale
TDS Factor 0.40 to 1.00 (default 0.67)
Temperature Range 0 to 100.0 °C
Resolution & Accuracy 0.1 °C & ± 0.5 °C
Temperature Compensation Automatic / Manual (0 to 80 °C)
Cell Constant 0.1, 1.0, 10.0 (selectable)
Temperature Coefficient 2%/°C
No. of Calibration Points Up to 5 max. (1 per range)
Special Functions Auto Power-Off after 17 minutes;
Hold & Self-diagnostic messages
Power 4 'AAA' x 1.4V batteries

Order Information

Order No.

Item
EC-TDS6/03K EcoScan TDS 6 Palmtop TDS Meter with electrode (EC-CONSEN91B) & carrying kit set (EC-ECOCONKIT)
EC-CONSEN91B 2 Stainless Steel Rings Ultem-body Electrode with ATC, cell constant K=1.0, 16 x 110 mm, BNC connector, 1m cable length (for CON 6/TDS 6/SALT 6)

 Sales Enquiry
 Service Enquiry

Products > By Parameter > TDS


Search Our Site
Product Search



Related Information

 Worldwide Offices

 Online Enquiry

 

 

 

 

 

Thermo Fisher Scientific Inc.

  Sitemap   Disclaimer   Privacy Policy

Copyright (c) 2005. Eutech Instruments Pte Ltd. All rights reserved. Powered by Xirlynx Innovations