Product
Specification |
Description |
Dissolved Oxygen |
Range |
0.00 to 20.00 mg/L ; 0.0 to 200.0 % |
Resolution |
0.01 mg/L ; 0.10 % |
Accuracy |
±1.5 % of full scale reading |
Temperature |
Range |
-9.9 to 125.0 ºC |
Resolution |
0.1 ºC |
Accuracy |
±0.5 ºC |
Sensor |
Pt100 ; 3 wire |
Compensation |
Temperature
Compensation |
Auto/manual |
Salinity
Compensation |
0.0 to 50.0 ppt
(manual setting and automatic correction) |
Pressure Compensation |
kPa/mmHg
(manual setting and automatic correction) |
Set
Point And Controller Functions |
Set Point 1 (SP1) /
Set Point 2 (SP2) |
0.00 to 20.00 mg/L or 0.0 to 200.0 % |
Switching DO Hysteresis |
0.1 to 1.0 mg/L or 0 to 10.0 % |
Function (Switchable) |
P control (pulse length/pulse frequency);
limit control |
Adjustable Period With Pulse Length Controller |
0.5 to 20 sec |
Adjustable Period With Pulse Frequency Controller |
60 to 120 pulse/min |
Pickup/Dropout Delay |
0 to 2000 sec |
Contact Outputs |
3 SPDT relays |
Switching Voltage/
Current/Power |
Max. 250 VAC / max. 3 A / max. 600 VA |
Alarm
Functions |
Function
(Switchable) |
Steady or fleet (pulse) |
Pickup
Delay |
0 to 2000 sec |
Switching
Voltage/
Current/Power |
Max. 250 VAC / max. 3 A / max. 600 VA |
Electrical
Data And Connections |
Transmitter Function |
0/4 to 20 mA scalable outputs for Dissolved Oxygen, galvanically isolated |
Hold Function Switch |
To freeze output current and deactivate control relays |
Load |
Max. 600 Ohm |
DO Input |
2-pin terminal |
Connection Terminal |
5-pole, 17-pole terminal, detachable blocks |
Display |
LCD |
7 segments display with symbols for status information |
Power Supply |
Input |
110 VAC
(jumper selectable) ;
48 to 62 Hz ; max. 7 VA
(For TSDOCTP1001) |
220 VAC
(jumper selectable) ;
48 to 62 Hz ; max. 7 VA
(For TSDOCTP1002) |
Main Fuse |
Slow-blow 250 V / 100 mA |
Pollution Degree |
2 |
Transient Overvoltage Category |
II |
EMC
Specifications |
Emitted Interference |
According to EN 50081-1 |
Immunity
To Interference |
According to EN 50082-1 |
Environmental
Conditions |
Operating Temperature Range |
-10 to 50 ºC |
Max. Relative Humidity |
80 % up to 31 ºC decreasing linearly to 50 %
at 40 ºC |
Mechanical
Specifications |
Dimensions (WxHxD) |
96 x 96 x 175 mm |
Weight |
700 g (unit) / 800 g (packed) |
Ingress Protection |
IP54 (front panel) |