Product
Specification |
Description |
Conductivity |
Range |
... to 10.00 ppm ; ... to 100.0 ppm ;
... to 1000 ppm ; ... to 5.00 ppt ;
... to 10.00 ppt ; ... to 100.0 ppt |
Resolution |
0.01 ppm ; 0.1 ppm ;
1 ppm ; 0.01 ppt ;
0.1 ppt |
Cell Constant |
0.1 ; 1 ; 10 |
Relative Accuracy |
±1 % of full scale |
Temperature |
Range |
-10 to 110 ºC |
Resolution |
0.1 ºC |
Accuracy |
±0.5 ºC |
Sensor |
Pt100; 2 or 3 wire |
Compensation |
Auto/manual (0 to 50 ºC) |
Coefficient |
0.0 to 5.0 % |
Normalisation |
10.0 ºC to 30.0 ºC |
TDS |
Factor |
0.40 to 1.0 |
Set
Point And Controller Functions |
Set Point 1 (SP1)/
set Point 2 (SP2) |
... to 10.00 ppm or ... to 100.0 ppm or
... to 1000 ppm or ... to 5.00 ppt or
... to 10.00 ppt or ... to 100.0 ppt |
Switching Conductivity Hysteresis |
0 to 10 % of full scale |
Switching TDS Hysteresis |
0 to 10 % of full scale |
Function |
Limit control |
Contact Outputs |
2 SPDT relays |
Switching
Voltage/
Current/Power |
Max. 250 VAC / max. 3 A / max. 600 VA |
Electrical
Data and Connections |
Transmitter Function |
4 to 20 mA scalable outputs for Conductivity, galvanically isolated |
Load |
200 Ohm |
Conductivity/TDS Input |
2-pin terminal |
Connection Terminal |
3-pin, 7-pin & 9-pin terminal, detachable blocks, 1 fixed connector |
Display |
LED |
7 segments display |
Power Supply |
Input |
85 to 250 VAC, 50/60 Hz |
Main Fuse |
250 mA, anti-surge, S504 Bussman |
Rated Input Current (Max.) |
100 mA |
Insulation Category |
II |
EMC Specifications |
Emitted Interference |
According to EN 61326 |
Immunity To Interference |
According to EN 61326 |
Environmental
Conditions |
Operating Temperature Range |
-10 to 50 ºC |
Max. Relative Humidity |
10 to 95 % (non-condensing) |
Mechanical
Specifications |
Dimensions (WxHxD) |
96 x 48 x 150 mm |
Weight |
300 g (unit) / 350 g (packed) |
Ingress Protection |
IP50 |