Product
Specification |
Description |
| pH
Range |
-2.00
to 16.00 pH |
| Resolution
& Accuracy |
0.01
pH & ±0.01 pH |
| mV
Range |
-1000
to 1000 mV / 0 to 100% (for -100 to
100 mV) |
| Resolution
& Accuracy |
1
mV / 0.1% & ±1 mV / ±0.2% |
| Temperature
Range |
-10.0
to 110 °C (14.0 to 230.0 °F) |
| Resolution
& Relative Accuracy |
0.1
°C / °F &
±0.5 °C / ±1.0 °F |
| Temperature
Sensor |
Pt100/
Pt1000 (jumper selectable); 2 or 3 wire |
| Temperature
Compensation |
Auto
/ Manual
(Reference at 25°C) |
| Set
Point and Controller Functions |
| Integral
Time |
0
to 999.9 minutes |
| Switching
pH Hysteresis |
0.1
to 1 pH |
| Function
(Switchable) |
P/PI
control (Pulse Length / Pulse Frequency);
Limit Control; Off |
| Adjustable
Period with Pulse Length Controller |
0.5
to 20 seconds |
Adjustable
Period with Pulse Frequency Controller
|
60
to 120 pulses/min |
| Pickup
/ Dropout Delay |
0 to 2000 seconds |
| Wash
Cycle |
0.1
to 200.0 hours |
| Wash
Duration |
1
to 2000 seconds |
Contact
Outputs, Controller
|
1
SPDT, 3 SPST relays |
| Switching
Voltage / Current / Power |
Max 250 VAC / Max 3A / Max 600VA |
| Alarm
Functions |
Function
(Switchable)
|
Latching
/ pulse |
| Pickup
Delay |
0
to 2000 seconds |
| Display |
LCD
|
UV
coat, backlit 14 segments display with
symbols for status information |
| Backlight
|
On/Off
selectable with four levels of brightness
control |
| Electromagnetic
Compliance (EMC) Specifications |
|
Emitted Interference |
According to EN 61326 |
| Immunity
to Interference |
According
to EN 61326 |
|
Environmental
Conditions |
| Ambient
Temperature Operating Range |
0
to 40 °C |
| Maximum
Relative Humidity |
80%
up to 31 °C decreasing linearly to 50%
at 40 °C |
| Power
Supply |
Input
|
80
to 250 VAC/DC 50/60 Hz Approx. 10 VA |
Main
Fuse |
250 mA, Anti-surge,
S504 BUSSMAN |
| Pollution
Degree |
2 |
| Transient
Overvoltage Category |
II |
| Electrical
Data and Connections |
| Transmitter
Function |
Two 0/4 to 20 mA scalable outputs for
pH/ORP and temperature, galvanically
isolated |
| CU 22
Function |
22
mA current output |
| Isolated
Output Voltage |
+12V
±0.5 V (Max 50 mA) |
Hold
Function Switch |
To
freeze output current and deactivate
control relays |
| Load
|
Max
600 W |
| pH
Input |
BNC
(1012 impedance); Asymmetrical/
Symmetrical |
Connection
Terminal
(Wall Mount) |
3-pin,
8-pin, 9-pin and 13-pin terminal, detachable
blocks |
Connection
Terminal
(Panel Mount) |
3-pin,
9-pin and 19-pin terminal, detachable
blocks |
| Mechanical
Specifications |
| Dimensions
(W x H x D) |
Wall-mount
Panel-mount |
144
x 144 x 110 mm
96 X 96 X 175 mm |
| Weight |
Wall-mount
Panel-mount |
745g
(Unit) / 1100g (Packed)
950g (Unit) / 1100g (Packed) |
| Environmental
Rating |
Wall-mount
Panel-mount |
IP66
(NEMA 4X)
IP54 (front panel) |