Home
Products > By Type > Controllers & Monitors Series

αlpha-pH560 1/4 DIN Industrial Controllers


 

 

Product Features

NIST or USA pH BUFFER OPTIONS
αlpha pH 550 Monitor and 560 Controller allow user to select USA (4.01, 7.00 or 10.01) or NIST (4.01, 6.86 or 9.18) pH buffer standards during calibration.

CUSTOM DUAL DISPLAY LCD
Primary and secondary displays are available for simultaneous display of pH, ORP, Conductivity and Temperature values. Allows instantaneous viewing of measured readings and setup configurations.

EASY PUSH-BUTTON OPERATION
Simply calibrate or adjust setup configuration with just a few touches of the buttons. Intelligent program in alpha pH 550 Monitor and 560 Controller prompts respective buffers for easy calibration. Similarly, push-button offset adjustment makes calibration simple in the conductivity models.

TEMPERATURE COMPENSATION
Compensates variance in the measured readings due to temperature fluctuations, either Automatically (with Pt 100) or Manually (Process and Calibration temperatures can be set separately).

PANEL OR WALL MOUNT, IP65 WEATHERPROOF HOUSING
Meters can be mounted on the wall or in a control panel. In addition, the IP65 rated housing allows the meter to be installed outdoors or indoors. Mounting accessories are included for both types of mounting.

SYMMETRICAL MODE OF OPERATION
Unique mode available in the alpha pH 550 Monitor and 560 Controller allow the meters to be operated in environments with electrical interferences. Simply configure the meter to the symmetrical mode and attach a potential matching pin.

EASY TO USE CONTROLLER FUNCTIONS
Two switching contacts acting as set-point relays simplify your task of controlling various processes. Separately adjustable high and low set-point hysteresis ensure chatter-free signal at the set-points.

ADVANCED FEATURES FOR BETTER CONTROL
Ability to set a time-delay up to 2000 seconds eliminates unneeded control and prevents false alarms, making it ideal for industrial environments.

BACK-LIT UV LIGHT PROTECTED LCD
Back-lit display ensures easy reading even in the dark. UV light protection minimises adverse effect of exposure to sunlight.

Specifications

Product Specification

Description
pH range 0.00 to 14.00 pH
Resolution 0.01 pH
Accuracy 0.01 pH
Temperature -10 °C to 125 °C
Resolution 0.1 °C
Relative Accuracy ±0.5 °C
Sensor Pt100 / Pt1000 (3 wire / 2 wire) (selectable)
Temperature Compensation Auto/Manual (reference at 25 °C)
Calibration Point Up to 2 points
Calibration Buffers USA : 4.01, 7.00, 10.01;
NIST : 4.01, 6.86, 9.18
Input Asymmetrical/Symmetrical
Controller Function Limit Control
Relay Operating Parameters LO (Relay active when measured value < set point)
HI (Relay active when measured value > set point
Time Delay 0 to 2000 seconds
Switching pH Hysteresis ±1.0 pH
Contact Output 2 SPST relay
Switching Voltage/Current Power Max 250VAC/Max 3A/MAX 600 VA
Display Dual Display Custom LCD
Backlight On/Off selectable
Power Requirements 110 or 220 VAC to 9 V DC Adapter
Dimension 96 x 96 x 66 mm (3.78 x 3.78 x 2.6 inch)
Weight 210 g

Order Information

Order No.

Item
TSPHCP0560 αlpha pH 560 pH/ORP Controller with 2 relay outputs and 110 & 220VAC to 9VDC adapter

 Sales Enquiry
 Service Enquiry

Products > By Type > Controllers & Monitors Series

 


Search Our Site
Product Search



Related Information

 Worldwide Offices

 Online Enquiry

 

 

 

 

 

Thermo Fisher Scientific Inc.

  Sitemap   Disclaimer   Privacy Policy

Copyright (c) 2005. Eutech Instruments Pte Ltd. All rights reserved. Powered by Xirlynx Innovations