The Alpha pH 800 controllers/transmitters combines consistent performance and sophisticated control functions with user-friendly features. Meters come with seven preset buffer values for quick, accurate auto-calibration; electrode offset function allows direct reading corrections without needing to remove electrode from the control system. Alarm delay and individual set-point hystereses in limit control mode prevent chattering, false alarms, and uncecessary down time.
::
2-in-1 controllers/transmitters can be configured to measure either pH or OPR (mV or %)
::
Auto-calibration with choice of NIST or USA buffer selection. Electrode slope and offset
is displayed after each successful calibration
::
Symmetrical mode option for clear, uninterrupted pH readings in electronically noisy environment
::
Antimony mode option for use with antimony electrodes in applications that involves corrosive Hydrofluoric Acid
::
Option of Auto or Manual Temperature Compensation. Three-wire system compensates for cable-length resistance errors
::
Galvanically-isolated, scaleable 0/4 to 20 mA output for high-quality output on peripheral devices
Applications
General: Useful
for any batch or on-line type application
that requires accurate pH or ORP control.
Water Purification/Treatment:
Use for batch and on-line control of incoming
process water, rinse water treatment, recirculating
system and waste water treatment.
Industrial:
Ideal for chemical processing, food processing.
aquarium, pharmaceutical, hydroponics and
waste control industries.
Specifications
Product
Specification
Description
pH
Range
-2.00 to 16.00 pH
Resolution
0.01 pH
Accuracy
±0.01 pH
ORP
Range
-1000 to 1000 mV
Resolution
1 mV
Accuracy
±1 mV
Temperature
Range
-9.9 to 125.0 ºC
Resolution
0.1 ºC
Accuracy
±0.5 ºC
Sensor
Pt100 / Pt1000 (jumper selectable); 2 or 3 wire
Compensation
Auto/manual
Set
Point And Controller Functions
Set point 1 (SP1) /
set point 2 (SP2)
-2.00 to 16.00 pH or -1000 to 1000 mV
Switching pH Hysteresis
0.1 to 1 pH
Switching ORP Hysteresis
10 to 100 mV
Function (Switchable)
Limit control
Pickup/Dropout Delay
0 to 2000 sec
Contact Outputs
2 SPDT relays
Switching Voltage/
Current/Power
Max. 250 VAC / max. 3 A / max. 600 VA
Electrical
Data And Connections
Transmitter Function
0/4 to 20 mA scalable outputs for pH/ORP, galvanically isolated
Hold Function Switch
To freeze output current and deactivate control relays
Load
Max. 600 Ohm
pH/ORP Input
BNC (1012 impedance); asymmetrical/symmetrical
Connection Terminal
5-pole, 17-pole terminal, detachable blocks
Display
LCD
7 segments display with symbols for status information
Power Supply
Input
110 VAC
(jumper selectable) ;
48 to 62 Hz ;
max. 0.75 VA
(For TSCTP0801)
220 VAC
(jumper selectable) ;
48 to 62 Hz ;
max. 0.75 VA
(For TSCTP0802)
Main Fuse
Slow-blow 250 V / 100 mA
Pollution Degree
2
Transient Overvoltage Category
II
EMC Specifications
Emitted Interference
According to EN 50081-1
Immunity To Interference
According to EN 50082-1
Environmental
Conditions
Operating Temperature Range
-10 to 50 ºC
Max. Relative Humidity
80 % up to 31 ºC decreasing linearly to 50 %
at 40 ºC