Home Corporate Information What's New Our Products Support Contacts
 
E-Brochure
Instruction Manual
CE Certificate
Press Release
Standard Hand-held Meters

CyberScan CON100/200 Conductivity/TDS/RS232 Portable Meters

  • Multi-Point Push-button Calibration (up to 5 points) for easy, quick calibration with ±1% full scale accuracy
  • Auto-Ranging for Conductivity/TDS measurements for fast response and best resolution over wide measurement range
  • User-Customization of meter functions like temperature coefficient, normalization temperature, stability, auto-power off, communication settings, etc.
  • RS232 Communication interface for downloading data to a computer or serial printer via a cable
  • Complimentary Data Acquisition Software helps organize data to import into most popular spreadsheet programs for report generation
  • Stores and recalls 16 data sets with temperature readings
  • Automatic Temperature Compensation (ATC) for the highest accuracy under fluctuating temperatures
  • Adjustable TDS Factor for more accurate measurement
  • Ready Indicator prompts whenever reading stabilizes 
  • Hold Function momentarily freezes reading for easy viewing and recording
  • Auto-Power Off saves battery power after non-use
  • Self-Diagnostic with message codes for easy troubleshooting
  • Large Custom LCD provides optimum viewing with user-friendly icons
  • Power with batteries or optional AC adapter for use in the field or at your bench-top with built-in stand

Applications

General: Monitor dissolved solids or conductivity levels quickly and easily in laboratories, field, schools and educational environments. 

Industrial: Use for testing pollution controls, water treatment, and testing for water hardness. Also useful for checking cooling towers, boiler water, font solutions in printing operations, brines, swimming pools, whirlpools and rise water.

Agricultural: Use in aquariums, fish farms, hydroponics, and fertilizer/chemical concentrations.

Specifications of CyberScan CON100/200 Meters

Product Code

EC-CON100 EC-CON200
Conductivity Range 0 to 19.99, 199.9, 1999 µS/cm; 19.99, 199.9 mS/cm
TDS Range Not Available 0 to 9.9, 99.9, 999 ppm; 9.99, 19.99 ppt;
Max. 199.9 ppt (depending on factor setting)
Resolution & Accuracy 0.05% Full Scale & ±1% Full Scale + 1 digit
Temperature Range 0.0 to 100.0 °C
Resolution & Accuracy 0.1 °C & ±0.5 °C
TDS Factor 0.40 to 1.00 (selectable)
No. of Calibration Points 5 (1 per range)
Temperature Compensation Automatic / Manual (0 to 80 °C)
Temperature Coefficient 0.0 to 10.0%/°C
Normalization Temperature 20.0 or 25.0 °C
Cell Constant 0.1, 1.0, 10.0 (selectable)
Memory 16 data sets
Special Functions Auto-ranging; Averaging/Stability; Self-diagnostic; 
Hold & Auto-Power Off after 20 minutes
Print Output No Yes (RS232C)
Power 4 'AAA' x 1.4V batteries or 9VDC power adapter



Copyright © 1997 Eutech Instruments Pte Ltd. marketing@eutechinst.com