Economy
Palm-top Meters
EcoScan
CON 5/TDS 5 Conductivity/TDS Meters
- Multi-Point Push-button Calibration
(up to 3 points) for easy, quick calibration with ±1% full
scale accuracy
- Auto-Ranging for Conductivity/TDS
measurements for fast response and best resolution over wide
measurement range
- Adjustable TDS Factor for more accurate
measurement (for TDS 5 model only)
- Automatic Temperature Compensation (ATC)
for the optimum accuracy under fluctuating temperatures
- Hold Function momentarily
freezes reading for easy viewing and recording
- Auto-Power Off saves battery
power after non-use
- Self-Diagnostic with message
codes for easy troubleshooting
- Large Custom LCD provides
optimum viewing even at a distance
- Protective Rubber Boot shields
meter from drops and features a sturdy built-in stand for easy bench-top
operation
- Meter Kit - contains everything
you need for calibration and measurement in a rugged carrying case
Applications
Routine Testing: For quick conductivity
or TDS checks in laboratories, field and schools. Useful in hydroponics
and agricultural industries.
Industrial:
For checking metal finishing, cooling tower water, printing
fountain solutions, boiler water, brines, drilling mud, rise tanks,
ponds, pollution control, recirculating systems, waste water and industrial
process systems.
Water Quality Testing: For analyzing
water hard water, untreated water, drinking water, effluent water,
pool water and incoming process water. Ideal for all types of quality
assurance and water quality testing.
Specifications
of EcoScan CON5/TDS 5 Meters
| Product
Code |
EC-CON5 |
EC-TDS5 |
| Conductivity Range |
0 to 199.9, 1999 µS/cm;
19.99 mS/cm |
Not Available |
| Resolution & Accuracy |
0.1, 1 µS/cm & 0.01 mS/cm;
±1% Full Scale |
- |
| TDS Range |
Not Available |
0 to 99.9, 999 ppm; 9.99 ppt |
| Resolution & Accuracy |
- |
0.1, 1 ppm & 0.01 ppt;
±1% Full Scale |
| TDS Factor |
- |
0.50 to 0.85 (default 0.67) |
| Temperature Range |
0 to 100.0 °C |
| Resolution & Accuracy |
0.1 °C & ± 0.5 °C |
| Temperature Compensation |
Automatic
/ Manual (0 to 80 °C) |
| Temperature Coefficient |
2%/°C |
| No.
of Calibration Points |
1
per range (Up to 3 max.) |
| Special Functions |
Auto Power-Off after 17 minutes;
Hold & Self-diagnostic messages |
| Power |
4 'AAA' x 1.4V batteries |

 
Copyright © 1997 Eutech