| Waterproof
Hand-held Meters
CyberScan DO300 Waterproof Dissolved Oxygen Meter
- Waterproof and Dustproof
Housing - IP67 rated, ideal for rugged harsh conditions.
Plus it floats!
- Automatically Compensates Salinity
and Barometric Pressure after manual input for increased accuracy
- Independent 100% and Zero Adjustment
Calibrations give meter high accuracy across entire measurement
range
- No Meter Warm-up Required with
Galvanic Cell - takes measurements immediately
and produces quick response
- Displays Electrode Diagnosis
gives slope, zero offset and millivolt values to help determine
when to service or replace electrode
- Stability Indicator prompts
whenever reading stabilizes
- Hold Function momentarily
freezes reading for easy viewing and recording
- Auto-Power Off saves battery
power after non-use
- Self-Diagnostic with message
codes for easy troubleshooting
- Adjustable Backlit Display
for viewing readings conveniently in poorly lit places
- Large, Easy-to-Read Custom
LCD Display with
simultaneous readings provides
optimum viewing with user-friendly icons
- Rugged Carrying Case (Included)
provides handy on-site meter calibration and measurement
Applications
Aquacultural: Use to monitor oxygen
levels in catfish and shrimp farming; game stocking ponds; ornamental
fish tanks and ponds; and in other fish farming applications.
Industrial:
Ideal for checks the quality of plant water intake and discharge,
wastewater and water treatment, recirculating systems and industrial
process systems.
Environmental: Use to test water
quality, monitor health of aquatic ecosystems, survey surface and
ground water drinking supplies, and meet EPA regulations.
Educational: Ideal for quick,
accurate DO readings in laboratories and schools.
Specifications
of CyberScan DO300 Meter
| Product
Code |
EC-DOWP300 |
| Dissolved
Oxygen Range |
0
to 19.9 mg/l or ppm |
| Resolution
& Accuracy |
0.01
mg/l, 0.1 ppm & ±1.5% Full Scale |
| %
Saturation of Oxygen |
0.0
to 199.9 % |
| Resolution
& Accuracy |
0.1
% & ±1.5% Full Scale |
| Temperature
Range |
0.0
to 100.0 °C (32.0 to 212 °F) |
| Resolution
& Accuracy |
0.1
°C (0.1 °F) & ± 0.3 °C (0.5 °F) |
| Salinity
Correction |
0.0
to 50.0 ppt |
| Barometric
Pressure Correction |
500
to 1499 mm Hg or 66.6 to 199.9 kPA |
| Method |
Manual
input for automatic correction |
| No.
of DO Calibration Points |
Up to 2 points; 100% in air and/ or
0% in known solution |
| Temperature
Compensation |
Automatic
/ Manual (0 to 100 °C) |
| Back-lit
Display |
Yes |
| Memory |
50
data sets |
| Special
Functions |
Stability
Indicator; Self-diagnostic; Electrode characteristics;
Hold & Auto-Power Off after20 minutes |
| Power |
4
'AAA' x 1.4V batteries |
 
Copyright © 1997 Eutech