Wall Mount/Panel Mount pH/ORP Controller/Transmitter

Built for the rugged industrial environment, the Alpha pH 2000 transmitter comes with an option for symmetrical mode of operation, making it ideal for electrically noisy environments. IP65 NEMA 4X casing is weatherproof and corrosion-resistant, protecting meter against harsh elements.

  • Wide pH Range - Measures from -2.00 to 16.00 pH with two-decimal accuracy
  • Dual Mode Operation - ORP mode measures in mV or % concentration, with independent calibration modes
  • Smart Calibration - Quick, easy push-button calibration with auto-buffer recognition and electrode status display
  • Antimony Electrode Support - Accepts Antimony electrodes for applications involving Hydrofluoric (HF) Acid
  • Master-Slave Operation - Control and measure pH and ORP simultaneously by placing two controllers side by side
  • Intelligent Diagnostics - Controller alerts when meter or electrode requires maintenance with trouble-shooting prompts in plain language
  • Symmetrical Mode Option - Ideal for electrically noisy environments, using floating differential mode operation

Operation Modes

Asymmetrical Mode Works well in environments with little or no electrical noise. When there is electrical noise, the noise acts as a common signal and is picked up by both pH and reference electrodes. Since the reference electrode is grounded to the amplifier’s ground potential, electrical noise appears only on the pH electrode, causing reading fluctuations. DC source noise (e.g., from electroplating tanks) typically results in stable but incorrect values.

Symmetrical Mode Avoids reference electrode grounding by using floating differential mode. Electrical noise appears equally on pH and reference electrodes and is rejected by the operational amplifier. Requires an electrode with solution ground (potential matching) pin.

Applications

General: Useful for any batch or on-line type application that requires accurate pH or ORP control.

Water Treatment: Use for batch and on-line control of incoming process water, rinse water treatment, recirculating system and waste water treatment.

Industrial: Ideal for chemical processing, food processing, aquarium, pharmaceutical, hydroponics and waste control industries.

Regulatory: Hook to recorder to document data for regulatory compliance.

Specifications

Product SpecificationDescription
pH
Range-2.00 to 16.00 pH
Resolution0.01 pH
Accuracy±0.01 pH
ORP
Range-1000 to 1000 mV / 0 to 100%
Resolution1 mV / 0.1%
Accuracy±1 mV / ±0.2%
Temperature
Range-10.0 to 125.0 ºC / 14.0 to 257.0 ºF
Resolution0.1 ºC / 0.1 ºF
Accuracy±0.5 ºC / ±1.0 ºF
SensorPt100 / Pt1000 (jumper selectable); 2 or 3 wire
CompensationAuto/manual (independent process/CAL temperature)
Controller Functions
Set PointsSP1/SP2: -2.00 to 16.00 pH or -1000 to 1000 mV or 0 to 100%
pH Hysteresis0.1 to 1 pH
ORP Hysteresis10 to 100 mV or 1 to 10.0%
Control ModesP/PI control (pulse length/pulse frequency/proportional integral); limit control; off
Pulse Length Period0.5 to 20 sec
Pulse Frequency60 to 120 pulse/min
Integral Action Time0 to 999.9 min
DelayPickup/Dropout: 0 to 2000 sec
Wash Cycle/Duration0.1 to 200.0 hr / 1 to 2000 sec
Contact Outputs1 SPDT, 3 SPST relays
Switching Specs1/8 HP at 125 VAC (max. 0.74 A / max. 93 VA)
1/8 HP at 250 VAC (max. 0.37 A / max. 93 VA)
Alarm Functions
ModeSteady or fleet (pulse)
Pickup Delay0 to 2000 sec
Switching Specs1/8 HP at 125 VAC (max. 0.74 A / max. 93 VA)
1/8 HP at 250 VAC (max. 0.37 A / max. 93 VA)
Electrical & Connections
Transmitter OutputTwo 0/4 to 20 mA scalable outputs for pH/ORP and temperature, galvanically isolated
Current OutputCU 22: 22 mA
Isolated Output12 V ±0.5 V (max. 50 mA)
Hold FunctionSwitch to freeze output current and deactivate control relays
LoadMax. 600 Ohm
pH/ORP InputBNC (1012 impedance); asymmetrical/symmetrical
Connection TerminalWall Mount: 3-pin, 8-pin, 9-pin & 13-pin terminal, detachable blocks
Panel Mount: 3-pin, 9-pin & 19-pin terminal, detachable blocks
Display
LCDUV coat, backlit 14 segments display with symbols
BacklightOn/off selectable with four levels of brightness
Power Supply
Input80 to 250 VAC/DC; 50/60 Hz; 10 VA
Main Fuse250 mA, anti-surge, S504 Bussmann
Pollution Degree2
Transient Overvoltage CategoryII
EMC Specifications
Emitted InterferenceAccording to EN 61326
Immunity To InterferenceAccording to EN 61326
Environmental
Operating Temperature0 to 40 ºC
Max. Relative Humidity80% up to 31 ºC decreasing linearly to 50% at 40 ºC
Mechanical
Dimensions (WxHxD)Wall Mount: 144 x 144 x 110 mm
Panel Mount: 96 x 96 x 175 mm
WeightWall Mount: 745 g (unit) / 1100 g (packed)
Panel Mount: 550 g (unit) / 950 g (packed)
Protection RatingWall Mount: IP66 (NEMA 4X)
Panel Mount: IP54 (front panel)

Order Information

Order No.Item
TSPHCTP2000WAlpha pH 2000 wall-mount pH/ORP controller/transmitter (unit only). Includes elbow BNC connector, terminal blocks, cable glands, gasket, etc
TSPHCTP2000PAlpha pH 2000 panel-mount pH/ORP controller/transmitter (unit only). Includes elbow BNC connector, terminal blocks, gasket, threaded rods, catch, etc