Product Features
- 5-Point Push-Button Calibration - with Auto-Buffer Recognition for quick calibration with no calibration mistakes
- Choice of pH Buffer Sets - USA, NIST, DIN or Low Ionic to cater wider applications
- Automatic Temperature Compensation (ATC) - for the highest accuracy under fluctuating temperatures
- Data Storage - Stores/ Recalls up to 100 Data Sets with temperature readings
- RS232 Output - for data logging and integration with other equipment
- Electrode Diagnostics - recall electrode slope, offset value and previous calibration values for easy reference
- Stability Indicator - prompts whenever reading stabilizes
- Hold Function - momentarily freezes reading for easy viewing and recording
- Auto-Power Off - conserves battery power and saves money
- Self-Diagnostic - with message codes for easy troubleshooting
- Large Custom LCD Display - with simultaneous readings provides optimum viewing with user-friendly icons
- Flexible Power Options - use with batteries or optional AC adapter for field or bench-top use with built-in stand
- Rugged Carrying Case - (Optional) provides handy on-site meter calibration and measurement
Applications
Industrial: For checks in water conditioning plants, cooling towers, food processing water testing, printing and chemical industries, and water/wastewater treatment. Use with ORP electrode to monitor plating operations, and other redox measurements.
Environmental/Agricultural: Use in ecology studies, aquariums, and hydroponics.
Educational: Ideal for quick, accurate routine pH checks in laboratories and schools.
Specifications
| Product Specification | Description |
|---|---|
| pH Range | -2.00 to 16.00 pH |
| Resolution & Accuracy | 0.01 pH ±0.01 pH |
| mV Range | ±199.9 mV (extended) & ±1999 mV |
| Resolution & Accuracy | 0.1 mV ±0.2 mV (within ±199.9 mV) 1 mV ±2 mV (beyond ±200 mV) |
| Temperature Range | -10.0 to 110.0 °C |
| Resolution & Accuracy | 0.1 °C ±0.5 °C |
| Temperature Compensation | Automatic / Manual (0 to 100 °C) |
| No. of Calibration Points | Up to 5 points |
| pH Buffer Options | USA, NIST |
| Auto-Hold | No |
| Memory | 50 data sets |
| Output | No |
| pH Slope & Offset Display | Yes |
| Special Functions | Stability Indicator; Self-diagnostic; Hold & Auto-Power Off after 20 minutes |
| Power | 4 ‘AAA’ x 1.4V batteries or 9VDC power adapter |
Order Information
| Order No. | Item |
|---|---|
| ECPH1102K | CyberScan pH 11 Portable pH/mV Meter with pH electrode, temperature probe, 2 electrode holders & CyberScan pH carry kit set (ECPHWPKIT) |
| ECPHWPKIT | CyberScan pH Carry Kit Set – CyberScan Plastic Carry Case comprises 1 x pH 4.01, pH 7.00, pH 10.01 buffer solutions (60 ml each) and 1 x empty rinse water bottle |