Product Features
- High Resolution and Accuracy - Up to ±0.002 pH Accuracy for precise measurement
- Advanced Display - Backlit Two Line, Dot Matrix LCD Display provides clear instructions for setup, measurement and calibration
- Multi-Point Calibration - Up to 5-point push-button calibration with Auto-Buffer Recognition for optimum full-range accuracy
- User-Customization - Enhanced flexibility for pH resolution and electrode diagnosis - pH slope/offset/mV base value, and previous calibration points
- Calibration Due Alarm - Prompts at user-defined interval to ensure regularly scheduled recalibration for ISO and GLP compliance
- Automatic Data Logging - Automatic downloading of data into memory or computer via cable at desired time intervals
- Expanded Memory - Stores and recalls up to 100 data sets either manually or on programmable real-time intervals
- Built-In Real-time Clock - Stamps stored data and calibration data with time & date for GLP compliance
- High/Low Set Point Alarm - Alerts once reading outside preset limits
- RS232 Communication - Interface to download data to printer or PC with complimentary Windows®-based DAS software
- Analogue Recorder - For external chart recording devices
- Temperature Compensation - Automatic Temperature Compensation (ATC) for best accuracy under fluctuating temperatures
- Stability Options - Multiple stability criteria options from instantaneous to moderate speed
- Quick Reference - Slide-out instruction card for quick reference
- Electrode Organization - Detachable electrode stand organizes multiple electrodes neatly

Display

to Computer
Applications
Laboratory/Industrial: Use for water quality testing, sewage treatment, soil/fertilizer testing, food processing (for HACCP compliance), geological testing. Great for R&D and QC applications in the food and pharmaceutical industries, paper/pulp processing, chemical industries and related fields. Suitable for most ASTM and EPA test procedures.
Educational: Perfect for research labs, suitable for all types of chemistry labs and ecology studies.
ORP: Use ORP electrode to monitor plating operations, ozonation system efficiency, and other redox measurements.
Specifications
Product Specification | Description |
---|---|
pH | |
Range | -1.000 to 14.000 pH |
Resolution & Accuracy | 0.01/0.001 pH & ±0.01/0.002 pH |
mV | |
Range | -1850 to +1850 mV |
Resolution & Accuracy | 0.1 mV & ±0.2 mV |
Temperature | |
Range | -5.0 to 105.0 °C (23.0 to 221.0 °F) |
Resolution & Accuracy | 0.1 °C (0.1 °F) & ±0.5 °C (0.9 °F) |
Calibration | |
No. of Points | 1 to 5 |
Buffer Options | USA: 1.68, 4.01, 7.00, 10.01, 12.45 NIST: 1.68, 4.01, 6.86, 9.18, 12.45 User 1: 1.68, 4.01, 6.86, 10.01, 12.45 User 2: 1.68, 4.01, 6.86, 10.01, 12.45 Custom: min. of 2 and max. of 5 points |
Other Features | |
Memory | 100 Data Sets with Date/Time stamping |
Temperature Compensation | Automatic/Manual (0 to 100 °C) |
pH Slope & Offset Display | Yes |
Print Output | Bi-directional RS232C, Recorder (±1850 mV) |
Display | Alpha-numeric Dot-matrix LCD with backlit |
Special Functions | Auto-buffer recognition, Date/time indexed (GLP compliance); Calibration due prompt, Alarm; Stability criteria Stability Indicator & Self-diagnostic |
Power | 9VDC, using AC adapters 110VAC or 220VAC, 50/60 Hz |
Order Information
Order No. | VAC | Item |
---|---|---|
ECPH110011S | 110 | CyberScan pH 1100 Bench pH/mV RS232 Meter with DAS, RS232C cable, ATC probe, integral electrode holder & power adapter |
ECPH110021S | 220 | CyberScan pH 1100 Bench pH/mV RS232 Meter with DAS, RS232C cable, ATC probe, integral electrode holder & power adapter |