Product
Specification |
Description |
| Conductivity |
| Range |
... to 1.999 µS/cm ; ... to 19.99 µS/cm ;
... to 199.9 µS/cm ; ... to 1999 µS/cm ;
... to 5000 µS/cm ; ... to 19.99 mS/cm ;
... to 199.9 mS/cm |
| Resolution |
0.001 µS/cm ; 0.01 µS/cm ;
0.1 µS/cm ; 1 µS/cm ;
5 µS/cm ; 0.01 mS/cm ;
0.1 mS/cm |
| Accuracy |
±1 % full scale reading |
| Cell Constant |
0.01 ; 0.1 ; 1 ; 10 |
| Temperature |
| Range |
-9.9 to 125 ºC |
| Resolution |
0.1 ºC |
| Accuracy |
±0.5 ºC |
| Sensor |
Pt100 / Pt1000 (jumper selectable) |
| Compensation |
Auto/manual (normalized at 25 ºC) |
| Coefficient |
Ultrapure water or linear 0.00 to 10.00 % |
| Set
Point and Controller Functions |
Set Point 1 (SP1) /
Set Point 2 (SP2) |
... to 1.999 µS/cm or ... to 19.99 µS/cm or
... to 199.9 µS/cm or ... to 5000 µS/cm or
... to 19.99 mS/cm or ... to 199.9 mS/cm |
| Switching Conductivity Hysteresis |
0 to 10 % of full scale |
|
Function (Switchable) |
P control (pulse length/pulse frequency); limit control |
Adjustable Period With Pulse Length Controller
|
0.5 to 20 sec |
| Adjustable Period With Pulse Frequency Controller |
60 to 120 pulse/min |
| Pickup/Dropout Delay |
0 to 2000 sec |
| Contact Outputs |
3 SPDT relays |
Switching Voltage/
Current/Power |
Max. 250 VAC / max. 3 A / max. 600 VA |
| Alarm
Functions |
| Function
(Switchable) |
Steady or fleet (pulse) |
Wash Cycle
|
0.1 to 199.9 hr |
| Wash Duration |
1 to 1999 sec |
| Pickup Delay |
0 to 2000 sec |
Switching Voltage/
Current/Power |
Max. 250 VAC / max. 3 A / max. 600 VA |
| Electrical
Data and Connections |
| Transmitter function |
0/4 to 20 mA scalable outputs for Conductivity,
galvanically isolated |
| Hold function switch |
To freeze output current and deactivate control relays |
| Load |
Max. 600 Ohm |
| Conductivity input |
2-pin terminal |
| Connection terminal |
5-pole, 17-pole terminal, detachable blocks |
| Power Supply |
| Input |
110 VAC
(jumper selectable) ;
48 to 62 Hz ; max. 7 VA
(For TSCONCTP1001)
|
220 VAC
(jumper selectable) ;
48 to 62 Hz ; max. 7 VA
(For TSCONCTP1002) |
| Main Fuse |
Slow-blow 250 V / 100 mA |
| Pollution Degree |
2 |
| Transient Overvoltage Category |
II |
| EMC
Specifications |
| Emitted Interference |
According to EN 50081-1 |
| Immunity
to Interference |
According to EN 50082-1 |
| Environmental
Conditions |
| Operating temperature range |
-10 to 50 ºC |
| Max. relative humidity |
80 % up to 31 ºC decreasing linearly to 50 % at 40 ºC |
| Mechanical
Specifications |
| Dimensions (WxHxD) |
96 x 96 x 175 mm |
| Weight |
700 g (unit) / 800 g (packed) |
| Ingress protection |
IP54 (front panel) |